6–2–81 Time Keeper RAM (for MVME167–002B) (FRU No. 613C)
Note
The MVME167–002B is supplied with the optional DICOM option.
Time Required
1 Hour
Tool Required
Screwdriver
IC Remover
Procedure
Refer to ILLUSTRATION 6–116 on page 6–205.
1.
Turn OFF the system.
2.
Remove the Side Cover Left (FRU 301). Refer to 6–2–31 on page 6–127.
3.
Unscrew the four Nest Cover screws.
4.
Remove the Nest Cover.
5.
Remove the CPU–A Assy (FRU 610). Refer to 6–2–78 on page 6–198.
6.
Remove the Time Keeper RAM on the SBC–A Assy (MVME167–002B).
Note
After replacing a time keeper RAM with new one, it is necessary to install the system ID number into
the time keeper RAM and to set up the factory preset data.