LOGIQ 400 SERVICE MANUAL
GE MEDICAL SYSTEMS
2127661
RENEWAL PARTS
6–143
REV 9
6–2–46 Front Caster Assy (FRU No. 318)
Time Required
15 Minutes
Tool Required
Screwdriver
Hexagonal Wrench
Procedure
Refer to ILLUSTRATION 6–69.
1.
Turn OFF the system.
2.
Put the system on a box.
3.
Unscrew four hexagonal bolts ( 1 – 4 ).
4.
Remove the Front Caster Assy.
FRONT CASTER ASSY DISASSEMBLY
ILLUSTRATION 6–69