6–4–5 Final Procedures
1.
Turn the system power ON.
2.
Check that the version number of the system software is correct, which is displayed on the CRT monitor screen
during the power up sequence.
3.
Set up the system preset parameters without regard to current settings. For set-up procedures, refer to
LOGIQ
400 Proprietary Manual, 2127662.
3–1 Set the Country/Zone properly. It takes approximately 40 minutes to complete this step.
3–2. Set other three parameters, Video Signal, Power Frequency and Modem Kind.
3–3. Reset the system using System Reset function of the Service Menu. Refer to LOGIQ
400 Proprietary
Manual, 2127662.
4.
Perform functional checks to verify that the system is functioning properly. Refer to Chapter 4, FUNCTIONAL
CHECKS, in this manual.
5.
Load the back-up user data into the system. For the procedure of loading back-up user data, refer to sub-section
4–3–3, Utility Menu in Chapter 4, FUNCTIONAL CHECKS.
6.
This completes the installation of the LOGIQ
400 system software.