![GE LOGIQ 400 Service Manual Download Page 316](http://html.mh-extra.com/html/ge/logiq-400/logiq-400_service-manual_128623316.webp)
GE MEDICAL SYSTEMS
2127661
LOGIQ 400 SERVICE MANUAL
RENEWAL PARTS
6–186
REV 9
6–2–73 MODD (FRU No. 520), MODD Fan (FRU No. 521), MODD Holder Assy (FRU No. 522) (continued)
Note
Make sure that the DIP switches and jumper switches located on the MODD are set as follows before
you install a new MODD onto the system.
ILLUSTRATION 6–100 shows the settings of jumper switches when the 3.5–inch HDD is installed on
the console.
Jumper Switches for another
type of MODD (2187012)
MODD (U0052ZA)
SETTINGS OF SWITCHES ON MODD
ILLUSTRATION 6–100
Note
If the 2.5–inch HDD is installed on the console, set the jumper switches as shown in
ILLUSTRATION 6–101.
MODD (U0052ZA)
Jumper Switches for another
type of MODD (2187012)
SETTINGS OF JUMPER SWITCHES
ILLUSTRATION 6–101