GE MEDICAL SYSTEMS
2127661
LOGIQ 400 SERVICE MANUAL
TM
8–62
OPTIONS
REV 5
8–5–8 Operational Check–out (Continued)
27. Select the ECG Wave section using the
or
key of Sub Menu Select keys.
28. Press
key to exit the ECG mode. The ECG waveform will disappear.
29. Connect the PCG sensor with the ECG panel. Refer to ILLUSTRATION 8–71.
Note
If your customer does not need the PCG sensor or you have not any PCG sensors, skip to step 42.
CONNECTING PCG SENSOR
ILLUSTRATION 8–71