6–2–61 Rear CONN Panel Assy (FRU No. 501) (continued)
4.
Unscrew three screws ( 1 – 3 ).
5.
Loosen one screw ( 4 ).
6.
Disconnect one flat cable ( 5 ) out of the CNTIF Assy or CNTIF2 Assy.
Note
The CNTIF2 Assy includes the Ethernet port for the DICOM option.
7.
Disconnect one flat cable ( 6 ) out of the AVIF Assy or BVIF Assy.
8.
Disconnect two connectors ( 7 and 8 ). One of them is connected with the Power Supply Unit, and another is
connected with the B/W Printer Panel.
9.
Remove the Rear CONN Panel Assy.