![GE LOGIQ 400 Service Manual Download Page 492](http://html.mh-extra.com/html/ge/logiq-400/logiq-400_service-manual_128623492.webp)
GE MEDICAL SYSTEMS
2127661
LOGIQ 400 SERVICE MANUAL
TM
8–54
OPTIONS
REV 5
8–5–6 Installing ECG Board Assy (Continued)
4.
Install the ECG Board Assy by screwing two screws ( 2 and 3 ) as shown in ILLUSTRATION 8–59.
5.
Connect the cable ( 1 : CB22) from the Mother Board with the connector on the ECG board assy. Refer to
ILLUSTRATION 8–59.
ECG BOARD ASSY INSTALLATION
ILLUSTRATION 8–59