LOGIQ 400 SERVICE MANUAL
GE MEDICAL SYSTEMS
2127661
INSTALLATION
2–5
REV 5
2–1–4 Recommended Ultrasound Room Layout
Table 2–1 provides the requirements for an ultrasound room:
TABLE 2–1
ULTRASOUND ROOM REQUIREMENTS
CURRENT RATING
15A (115V, 100V) ; 7.5A (220–240V) CIRCUIT BREAKER
RADIATION SHIELDING
NONE REQUIRED for ULTRASOUND ENERGY
TEMPERATURE
20–26 DEG. C (68–79 DEG F) for PATIENT COMFORT
HUMIDITY
50% to 70% for PATIENT COMFORT
HEAT DISSIPATION
2000 BTU/Hr for LOGIQ
400 ;
FLOOR LOADING
Approximately 680 – 800 kg/m
2
without Accessories
FLOOR CONDITION
Gradient : WITHIN 5 degrees
LOGIQ
400 Weight
145 kg (320lbs) without Accessories
POWER SOURCE
220–240VAC, 50Hz, SINGLE PHASE For Europe Version
115VAC, 60Hz, SINGLE PHASE
For USA Version