![GE LOGIQ 400 Service Manual Download Page 393](http://html.mh-extra.com/html/ge/logiq-400/logiq-400_service-manual_128623393.webp)
GE MEDICAL SYSTEMS
2127661
LOGIQ 400 SERVICE MANUAL
RENEWAL PARTS
6–263
REV 9
6–6 REPLACING 2.5–INCH HDD ASSY WITH 3.5–INCH HDD ASSY
6–6–1 Introduction
This section provides the installation procedures of the 3.5–Inch HDD Assy using the Installation Kit for LOGIQ
400
console.
6–6–2 Time Required
1 hour
6–6–3 Parts Required
ITEM
PART NAME
QTY
PART #
REMARKS
1.
3.5–inch HDD Assy
2
Part number is different
depending on the software
version.
2.
HDD Change Kit
1
2160057
HDD Holder2 Assy
1
2145973
For 3.5–inch HDD Assy,
including screws
CB24N Cable Assy
1
2145889
MOD and HDD to MOTHER
CB5N Cable Assy
1
2145890
HDD to MOTHER
CB45N Cable Assy
1
2145891
HDD to HDD LED Ass
through CB46
Tie Wrap
2
This part is required only for
LOGIQ
500
Note
If the console includes the Quantum
FireBall
1280S 3.5–inch HDD Assy, it is necessary to use the
CB45N Cable Assy for the HDD LED Assy. Because the HDD LED Assy is not available on the con-
sole with this type of 3.5–inch HDD Assy.
6–6–4 Procedures
(A) Removing 2.5–inch HDD Assy, HDDB Assy and Cables
1.
Turn the main power switch OFF.
2.
Wait a while until the power off process is completed.
3.
Set the circuit breaker located at the rear of Power Supply unit to OFF position (lower side).
4.
Unplug the unit. Follow the OSHA Lockout/Tagout Requirements (29 CFR 1910.147, Direction 46–015902).
5.
Remove the Keyboard Panel Assy. Refer to 6–2–49 on page 6–147.
6.
Remove the Front Base Cover. Refer to 6–2–35 on page 6–131.
7.
Remove the Side Cover Right. Refer to 6–2–32 on page 6–128.