![GE LOGIQ 400 Service Manual Download Page 423](http://html.mh-extra.com/html/ge/logiq-400/logiq-400_service-manual_128623423.webp)
GE MEDICAL SYSTEMS
2127661
LOGIQ 400 SERVICE MANUAL
7–5
PERIODIC MAINTENANCE
REV 2
7–2–2
Cleaning (Continued)
b. Clean the Fans and the Finger Guards by using a vacuum cleaner and bristled brush.
c. Clean two Air Filters and one PW Air Filter by using a bristled brush. If these filters are damaged, replace
them. For the positions of three Air Filters, refer to ILLUSTRATION 7–2.
AIR FILTERS REMOVAL
ILLUSTRATION 7–2
d. Clean out the interior of the Console by using a vacuum cleaner.