LOGIQ 400 SERVICE MANUAL
GE MEDICAL SYSTEMS
2127661
INSTALLATION
2–3
REV 0
2–1 PREINSTALLATION
2–1–1 Introduction
This section describes various general electrical, operational, and environmental considerations that must be consid-
ered before installing the LOGIQ
400 Ultrasound unit.
2–1–2 Power Line Requirements
The following power line parameters should be monitored for one week before installation. We recommend that you
use an analyzer Dranetz Model 606–3 or Dranetz Model 626:
PARAMETER
:
LIMITS
Voltage Range
: Japan.
: 100 VAC
±
10% (90–110 VAC)
: Europe
: 220–240 VAC
±
10% (198–264 VAC)
USA
: 120 VAC
±
10% (108–132 VAC)
Power
: Japan
: MAX. 1000 VA
: Europe
: MAX. 1000 VA
: USA
: MAX. 1000VA
Line Frequency
: All applications
: 50/60Hz (
±
2Hz)
Power Transients
: Less than 25 % of nominal peak voltage for less than 1 millisecond for any type of
transient, including line frequency, synchronous, asynchronous, or aperiodic
transients.
Decaying Oscillation : Less than 15 % of peak voltage for less than 1 millisecond.