2127661
LOGIQ 400 SERVICE MANUAL
GE MEDICAL SYSTEMS
RENEWAL PARTS
6–224
REV 9
6–2–90 Power Inlet Unit (FRU No. 706 for 115V, FRU No. 707 for 220V)
Time Required
15 Minutes
Tool Required
Screwdriver
Procedure
Refer to ILLUSTRATION 6–125 on page 6–225.
1.
Turn OFF the system and unplug the main power cord.
2.
Remove the Rear Cover (FRU 303). Refer to 6–2–33 on page 6–129.
3.
Unscrew six screws ( 1 – 6 ).
4.
Pull the Power Inlet Unit out.
5.
Disconnect one connector ( 7 ) .
6.
Unscrew two screws ( 8 and 9 ) to remove two FG wires.
7.
Remove the Power Inlet Unit.