LOGIQ 400 SERVICE MANUAL
GE MEDICAL SYSTEMS
2127661
RENEWAL PARTS
6–144
REV 9
6–2–47 Rear Caster Assy (FRU No. 319)
Time Required
15 Minutes
Tool Required
Screwdriver
Hexagonal Wrench
Procedure
Refer to ILLUSTRATION 6–70.
1.
Turn OFF the system.
2.
Put the system on a box.
3.
Unscrew four screws ( 1 – 4 ).
4.
Remove the Rear Caster Assy.
REAR CASTER DISASSEMBLY
ILLUSTRATION 6–70