LOGIQ 400 SERVICE MANUAL
GE MEDICAL SYSTEMS
2127661
INSTALLATION
2–8
REV 0
2–2 INSTALLATION
2–2–1 Introduction
This section contains many of the procedures required to install the LOGIQ
400 console.
2–2–2 Average Installation Time
The LOGIQ
400 has been designed to be installed and checked out by an experienced service technician in approxi-
mately four hours. LOGIQ
400 consoles with optional equipment may take slightly longer.
2–2–3 Installation Warnings
1.
Since the LOGIQ
400 weighs approximately 320 lbs (145 kg) without options, preferably two people should
unpack it. Two people are also preferable for installing any additional bulky items.
2.
There are no operator serviceable components. To prevent shock, do not remove any covers or panels. Should
problems or malfunctions occur, unplug the power cord. Only qualified service personnel should carry out servic-
ing and troubleshooting.
Note
For information regarding packing labels, refer to ILLUSTRATION 2–3, LABELS ON PACKAGE.
2–2–4 Checking the Components
When a new system arrives, check that any components are not damaged and are not in short supply. If shipping
damage or shortage occurs, contact the address shown in Chapter 1.