GA091568, R
EVISION
5
VIVID E9 S
ERVICE
M
ANUAL
8 - 170
Section 8-6 - Top Console Parts Replacement
8-6-36
Up-Down Button Board (Buttons Frame UI Assy) replacement
These buttons are also called the XYZ Buttons.
The XYZ Buttons, located on the palm rest, control the XY and Z movement.
•
Push one of the buttons marked “1”, and then reposition the XY (horizontal) location of the upper
console. Three seconds after pushing the left button, the 4 motors in the XY Frog Leg lock to
stabilize the console.
•
Toggle one of the buttons marked “2” to reposition the Z (vertical) location of the upper console.
Toggle up to raise, or down to lower, the console.
Figure 8-177 Top Console adjustment controls
Figure 8-178 XYZ Buttons
WARNING
WARNING
DO NOT TOUCH ANY BOARDS WITH INTEGRATED CIRCUITS PRIOR TO TAKING THE
NECESSARY ESD PRECAUTIONS.
1. ALWAYS CONNECT YOURSELF, VIA AN ARM-WRIST STRAP, TO THE ADVISED
ESD CONNECTION POINT LOCATED ON THE REAR OF THE SCANNER (NEAR THE
POWER CONNECTOR).
2. FOLLOW GENERAL GUIDELINES FOR HANDLING OF ELECTROSTATIC SENSITIVE
EQUIPMENT.
1. Lock and brake release button: Unlock and move the Top Console horizontal.
2. Up/Down button: Move the Console up or down
Summary of Contents for GA000100
Page 1: ...GE Healthcare OPERATING DOCUMENTATION VIVID E9 Service Manual Part Number GA091568 Revision 5 ...
Page 2: ......
Page 11: ...GE HEALTHCARE DIRECTION GA091568 REVISION 5 VIVID E9 SERVICE MANUAL ix ZH CN KO ...
Page 34: ...GE HEALTHCARE DIRECTION GA091568 REVISION 5 VIVID E9 SERVICE MANUAL xxxii ...
Page 800: ...GE HEALTHCARE DIRECTION GA091568 REVISION 5 VIVID E9 SERVICE MANUAL Index 12 ...
Page 801: ......