Chapter 8 - Replacement procedures
8 - 93
8-6-5-3
Install the LCD Cables
NOTE:
Do not over-stretch cables. Before putting covers on arms, be sure arm has full pivot motion/
rotation without stressing cables. If replacing more than one cable, begin with the thickest cable
first.
Follow these steps to install the LCD Cables:
1.) Holding the monitor-end of the cable and starting at the console-end of the first arm section, thread
the replacement cable(s) through the arm sections toward the monitor-end.
2.) Feed cable through opening opposite the arm lock mechanism side of first arm section.
3.) Make sure the cable markings
line up with the midpoint placement location for the first arm (Figure
Figure 8-102 Feed Cable through First Arm
Figure 8-103 First (left) and Second Arm Section Midpoint Locations
Console-End of Arm
Feed cables through
opening opposite the
Arm Lock Mechanism
side
Arm Lock knob
Arm Lock
Mechanism
Midpoint Locations
First Arm - Midpoint Location
Second Arm - Midpoint Location