GE H
EALTHCARE
D
IRECTION
GA091568, R
EVISION
5
VIVID E9 S
ERVICE
M
ANUAL
Chapter 8 - Replacement procedures
8 - 227
NOTE:
Even if only one Front Plane Board is replaced, remove both so that the Relay Board and GTX
Board(s) are returned to proper position before installing the Front Plane Board.
4.
Repeat the same also for the RX Board.
5.
Holding the upper and lower edges of the upper Front Plane Board with both hands, rock the upper Front Plane Board
evenly away from the GRLY and RX boards. Be careful not to bend the connector pins.
6.
Repear the previous step for the Lower Front Plane Board.