Chapter 8 - Replacement procedures
8 - 175
8-7-1-2
XY Mechanism removal procedure
1.) From the rear side of the VIVID E9, remove the four (4x) fixing screws.
2.) Remove the XY mechanism.
Figure 8-180 Operator Panels have been removed
Figure 8-181 Remove screws for the XY (Frogleg) mechanism
NOTE!
MAIN CABLE IS NOT SHOWN IN
THIS PICTURE.
REMOVE FOUR (4X) SCREWS