Chapter 8 - Replacement procedures
8 - 125
8-6-16-2
Remove the Operator Panel, Lower
Follow these steps to remove the Operator Panel:
1.) Pull the Alphanumeric keyboard to its extended position to get access to the screws in the next step.
2.) The screws are available from the underside of the Operator Panel tray. Unscrew and remove the
four screws used for fixing the Operator Panel assembly to the Operator Panel tray.
3.) Loosen the screws for the cables grounding on the Bulkhead Bracket and move the OP Cables
away from the bracket.
Figure 8-137 OP seen from below
Figure 8-138 OP Cables Grounding
1 - Cable, A/N Keybd.
2 - USB2
3 - USB1
4 - Bulkhead cable
5 - HDMI Cable
6 - OP Cable
1
2
3
4
5
6