GE H
EALTHCARE
D
IRECTION
GA091568, R
EVISION
5
VIVID E9 S
ERVICE
M
ANUAL
5 - 80
Section 5-10 - Power distribution
5-10-2-4
Bidirectional Signals
USB bus
5-10-2-5
Outputs
Table 5-23 USB connector
DESCRIPTION
COMMENTS
Originates on the BEP.
The USB is used to set the power supply to the correct
transmit (XD) voltages.
Table 5-24 Power outputs sheet 1 of 2
CONNECTOR
DESCRIPTION
COMMENTS
P2
115 VAC OUT ON TWO CONNECTORS
PERIPHERALS (B/W PRINTER)
P3
Card Rack Connector
FRONT END PROCESSOR
- The ‘TSON_STRB’ signal is a watchdog for the
Transmit Signal (TS).
P4
Card Rack Connector
FRONT END PROCESSOR
EXT_SYNC is a 200kHz clock coming from the
GFI board.