GE H
EALTHCARE
D
IRECTION
GA091568, R
EVISION
5
VIVID E9 S
ERVICE
M
ANUAL
8 - 66
Section 8-5 - Replacing Covers and Bumpers
8-5-15-3
Main Cable Cover installation procedure
Follow these steps to install the Main Cable Cover:
1.) Install the Main Cable Cover so the Column Cover Assembly overlaps Main Cable Cover edges
(see
).
2.) With the console raised to its full height, install the 3 screws to secure the Main Cable Cover and
Column Cover Assembly.
3.) Lower the console and install the lower right side screw (See:
Figure 8-62 "With console lowered,
lower right side screw placement" on page 8-63
4.) Position the lower Column Cover tab inside the Front Cover.
5.) Install the Top Cover.
6.) Install the Side Covers.
Figure 8-67 Position the column cover tab inside the front cover