GE H
EALTHCARE
D
IRECTION
GA091568, R
EVISION
5
VIVID E9 S
ERVICE
M
ANUAL
Chapter 8 - Replacement procedures
8 - 255
8-11-2-5
Install BEP5 Power Supply
Follow these steps to install the BEP5 Power Supply:
1.) With the 2 pins pointing to the back of the BEP, guide the pins into the 2 holes in the top right of the
BEP. Apply a pressure to make it slide into position.
2.) With the pins in place, replace the 2 screws that secure the front of the power supply to the BEP
frame.
3.) Install the BEP Cover as described in these steps:
a.) Insert the base of the BEP Cover inside the base of the BEP.
Be sure the bottom lip of the BEP Cover rests inside the BEP.
b.) Tilt the top of the BEP Cover toward the VIVID E9.
c.) Install the fixing screw(s) for the BEP Cover:
-
On BEP6 there are two finger screws on the top of the BEP Cover.
-
On newer BEP5 there are three Phillips screws, one on the top and one on each side of
the BEP Cover.
-
On older BEP5 there is one finger screw at the top of the BEP Cover.
4.) Replace the Left Side Cover.
NOTICE
Be careful not to pinch any of the cables when installing the BEP Cover.