Chapter 8 - Replacement procedures
8 - 99
8-6-7-2
Remove the Upper OP Panel/Touch Panel Assembly
Follow these steps to remove the Upper OP Panel/Touch Panel assembly. Ensure that the OP Panel is
in its uppermost position with the LCD out of the way:
1.) At the rear of the VIVID E9, release the console’s XY mechanism by inserting a screwdriver into the
release point and pressing until release. Pull the console out to its extended position to gain access
to the screws in the next step.
2.) Remove four screws with washers from the Operator Panel’s back cover.
Figure 8-108 Upper OP Panel//Touch Panel Assembly
Figure 8-109 XY / Frogleg mechanism release
Figure 8-110 Upper Operator Panel’s Back Cover
OPERATOR PANEL,
UPPER
XY RELEASE
FASTENING SCREWS, 2 ON EACH SIDE
(ONE SIDE ILLUSTRATED)