GE H
EALTHCARE
D
IRECTION
GA091568, R
EVISION
5
VIVID E9 S
ERVICE
M
ANUAL
Chapter 8 - Replacement procedures
8 - 249
8-11-1-5
Back End Processor (BEP) installation procedure
Follow these steps to install the BEP:
1.) Slide the BEP into the left side of the chassis frame.
2.) Install the two (2x) hex key screws at the inside base of the BEP.
3.) Reach inside the BEP and connect the Boundary Scan Cable to the BEP’s Backplane.
NOTE:
If you do not reconnect the internal BEP cable (Boundary Scan Cable) to the Backplane, the
VIVID E9 will power up, but will not scan. The BEP will launch into simulator mode.
4.) Install the BEP Cover as described in these steps:
a.) Insert the base of the BEP Cover inside the base of the BEP.
Be sure the bottom lip of the BEP Cover rests inside the BEP.
b.) Tilt the top of the BEP Cover toward the VIVID E9.
c.) Install the fixing screw(s) for the BEP Cover:
-
On BEP6 there are two finger screws on the top of the BEP Cover.
-
On newer BEP5 there are three Phillips screws, one on the top and one on each side of
the BEP Cover.
-
On older BEP5 there is one finger screw at the top of the BEP Cover.
5.) Replace the Printer Bracket at the top of the BEP.
Be sure the lip, on the underside of the bracket, hooks on the edge of the Card Rack, and the 3 tabs
insert into the slots on the top of the BEP frame. The lip “clamps” the Card Rack and BEP together.
This is a tight fit.
6.) Lower the Lower Column Cover.
7.) Position the Front Cover to engage the Column Cover Stop Tabs.
8.) Latch the two latches that clamp the Printer Bracket to the top of the BEP.
9.) Slide the B&W printer into the Printer Bracket and connect the cables to the back of the printer.
10.)Tighten the Printer Bracket Wing Nut to secure the printer.
11.)Connect all I/O cables at top and/or face of the BEP, including the printer cables.
12.)Install the Top Cover (if removed), the Front cover (if removed) and the side covers.
13.)If applicable, install the correct software on the new BEP.
14.)Enter the VIVID E9’s TCPIP settings and Option strings.
15.)Restore the Patient Archive images, Report Archive, User-Defined Configuration (Customer
Presets), and Service settings.
•
NOTICE
Be careful not to pinch any of the cables when installing the BEP Cover.
Summary of Contents for GA000100
Page 1: ...GE Healthcare OPERATING DOCUMENTATION VIVID E9 Service Manual Part Number GA091568 Revision 5 ...
Page 2: ......
Page 11: ...GE HEALTHCARE DIRECTION GA091568 REVISION 5 VIVID E9 SERVICE MANUAL ix ZH CN KO ...
Page 34: ...GE HEALTHCARE DIRECTION GA091568 REVISION 5 VIVID E9 SERVICE MANUAL xxxii ...
Page 800: ...GE HEALTHCARE DIRECTION GA091568 REVISION 5 VIVID E9 SERVICE MANUAL Index 12 ...
Page 801: ......