Chapter 4 - General procedures and Functional checks
4 - 43
4-3-6
Tissue Velocity Imaging (TVI) Checks
4-3-6-1
Introduction
TVI calculates and color codes the velocities in tissue. The tissue velocity information is acquired by
sampling of tissue Doppler velocity values at discrete points. The information is stored in a combined
format with grey scale imaging during one or several cardiac cycles with high temporal resolution.
4-3-6-2
Preparations
1.) Connect one of the probes, to the scanner’s left-most probe connector.
-
See
, page
for info about connecting the probes
2.) Turn ON the scanner
The 2D Mode window is displayed (default mode).
3.) If needed, adjust the Display’s Brightness and Contrast setting.
4.) Press
TVI
.
5.) Use the trackball (assigned function: Pos) to position the ROI frame over the area to be examined.
6.) Press
Select
. The instruction Size should be highlighted in the trackball status bar.
NOTE:
If the trackball control pointer is selected, press
trackball
to be able to select between Position
and Size controls.
7.) Use the trackball to adjust the dimension of the ROI.
4-3-6-3
Adjust the TVI Controls
•
To reduce quantification noise (variance), the Nyquist limit should be as low as possible, without
creating aliasing. To reduce the Nyquist limit: Reduce the
Scale
value.
NOTE:
The Scale value also affects the frame rate. There is a trade off between the frame rate and
quantification noise.
•
TVI provides velocity information only in the beam direction. The apical view typically provides the
best window since the beams are then approximately aligned to the longitudinal direction of the
myocardium (except near the apex). To obtain radial or circumferential tissue velocities, a
parasternal view must be used. However, from this window the beam cannot be aligned to the
muscle for all the parts of the ventricle.
NOTE:
PW will be optimized for Tissue Velocities when activated from inside TVI.
Summary of Contents for GA000100
Page 1: ...GE Healthcare OPERATING DOCUMENTATION VIVID E9 Service Manual Part Number GA091568 Revision 5 ...
Page 2: ......
Page 11: ...GE HEALTHCARE DIRECTION GA091568 REVISION 5 VIVID E9 SERVICE MANUAL ix ZH CN KO ...
Page 34: ...GE HEALTHCARE DIRECTION GA091568 REVISION 5 VIVID E9 SERVICE MANUAL xxxii ...
Page 800: ...GE HEALTHCARE DIRECTION GA091568 REVISION 5 VIVID E9 SERVICE MANUAL Index 12 ...
Page 801: ......