5 - 30
Section 5-8 - Front End Processor (FEP)
5-8-2-2
Phased and Linear Array probes
•
Phased and Linear Array probes consist of several identical transducer elements (e.g. 64, 128,
192).
•
Four probes can be connected to the system at the same time. The probe connectors are physically
located on the Relay Board, where one is selected and connected to the transmitter (GTX boards)
and receiver (GRX board) through a number of relays.
5-8-2-3
Receiver signal path
NOTE:
The number of cards depend on the VIVID E9 model, and options installed.
•
The reflected signal from body structures and blood cells are routed from the probe, via the Relay
Board and the Front Plane to the GRX (receiver) boards, where pre-amplification and Analog Time
Gain Compensation (ATGC) is performed. The gain is determined by an analog control signal
(ATGC) generated by the Global Radio Frequency Interface board (GFI).
•
The output channels from the GRX boards are fed to the DRX boards where the signals in each
receiver channel are A/D converted. Then the beamforming for the received signals takes place.
Figure 5-15 The Ultrasound Receiver
RELAY
BOARD
GRX
GFI
BOARD
PROBES
DOPPLER PROBE
XD_BUS
FE_BUS
DRX
GTX
STA_BUS
LVDS_BUS
DLP
DL
FE_BUS
CW_BUS (RX)
PCIe BUS
TO/FROM BEP
Summary of Contents for GA000100
Page 1: ...GE Healthcare OPERATING DOCUMENTATION VIVID E9 Service Manual Part Number GA091568 Revision 5 ...
Page 2: ......
Page 11: ...GE HEALTHCARE DIRECTION GA091568 REVISION 5 VIVID E9 SERVICE MANUAL ix ZH CN KO ...
Page 34: ...GE HEALTHCARE DIRECTION GA091568 REVISION 5 VIVID E9 SERVICE MANUAL xxxii ...
Page 800: ...GE HEALTHCARE DIRECTION GA091568 REVISION 5 VIVID E9 SERVICE MANUAL Index 12 ...
Page 801: ......