Chapter 5 - Components and functions (theory)
5 - 39
5-8-4-2
Location in the Unit
The Relay board is located in the Front End Rack on the end nearest to the front of the scanner.
5-8-4-3
Input DC Voltages:
•
+15VDC
•
+6VDC
•
-6VDC
•
-15VDC
•
+100V PMXVPP (Voltage for Probe MUX)
•
-100V PMXVNN (Voltage for Probe MUX)
Voltages provided by regulators on the board:
•
+12 VDC
•
+5 VDC
•
+3.3 VDC
•
-3.3 VDC
•
-5 VDC
and
•
LVDC (3.3 V/2.5A) for RT3D probe
5-8-4-4
Input Signals
•
Transmit period:
XMIT Pulses via the XD_BUS
•
Receive period:
Echo signals from the selected probe
5-8-4-5
Output Signals
•
Transmit period:
XMIT Pulses to the selected probe
•
Receive period:
Echo signals via XD_BUS to GRX cards