Section 8-5 - Replacing Covers and Bumpers
8-5-14-3
Column Cover Assembly Installation
Follow these steps to install the Column Cover Assembly
1.) Install the Main Cable Cover so the Column Cover Assembly overlaps Main Cable Cover edges.
2.) Install screws to Column Cover Assembly (tighten by hand).
3.) Position the lower Column Cover tab inside the Front Cover.
4.) Install the Top Cover.
5.) Install the Side Covers.
Figure 8-65 Position the column cover inside the front cover