Section 6-5 - Touch Screen Calibration
Section 6-5
Touch Screen Calibration
The
Touch Screen Calibration
is found on the
Operator Panel Test Dialog
.
Follow these steps to open the Operator Panel Test Dialog:
1.) Select
Config
(
F2
).
2.) Log on as
ADM
.
3.) Select
System > Test
.
4.) Select
Operator Panel Test
. This will open the Operator Panel Test Dialog (
).
The
Calibrate
button is used to start to calibrate the position of the touch sensitive area on the Touch
screen relative to the graphic artwork.
Follow these steps to calibrate the Touch screen:
1.) Select
Calibrate
. The Touch screen goes blank and display a cross in the screen’s upper left
corner.
2.) Point your finger on the centre of the cross and press slightly. This calibrates this point, and the
cross moves to the screen’s upper right corner.
3.) Point your finger on the centre of the cross and press slightly. This calibrates this point, and the
cross moves to the lower right corner.
4.) Point your finger on the centre of the cross and press slightly. This calibrates this point, and the
cross moves to the lower left corner.
5.) Point your finger on the centre of the cross and press slightly. This calibrates this last point.
6.) Select
Done
. The calibration is complete.
Figure 6-9 The Operator Panel Test Dialog
Summary of Contents for GA000100
Page 1: ...GE Healthcare OPERATING DOCUMENTATION VIVID E9 Service Manual Part Number GA091568 Revision 5 ...
Page 2: ......
Page 11: ...GE HEALTHCARE DIRECTION GA091568 REVISION 5 VIVID E9 SERVICE MANUAL ix ZH CN KO ...
Page 34: ...GE HEALTHCARE DIRECTION GA091568 REVISION 5 VIVID E9 SERVICE MANUAL xxxii ...
Page 800: ...GE HEALTHCARE DIRECTION GA091568 REVISION 5 VIVID E9 SERVICE MANUAL Index 12 ...
Page 801: ......