Chapter 8 - Replacement procedures
8 - 283
8-11-9-3
Patient I/O Interface Cable installation
Follow these steps to install the
Patient I/O Interface Cable
:
1.) Connect the two connectors (Power and USB) to the Patient I/O.
2.) Connect the other end of the cable to
CN USB11
on the BEP’s motherboard.
3.) Install the
BEP Power Board
.
4.) Install the
BEP Cover
.
5.) Install the
Left Side Cover
.
CAUTION
BE AWARE OF CORRECT ORIENTATION OF THE PATIENT I/O CABLE CONNECTOR WHEN
CONNECTING TO CN USB11 ON THE BEP'S MOTHERBOARD.
ON SOME BEPS, THE KEYING TO PREVENT ERROR MAY BE MISSING.