Configuration
2026419-033E
CIC Pro™
6-5
3.
Analyze the Check Centrals output:
a.
Identify time zone or daylight saving time network errors:
Verify that all the CIC Pro centers configured on the CARESCAPE
Network are listed.
NOTE
The target CIC Pro center running the Unity Time Zone Discovery
Tool utility is the baseline device used for determining time zone or
daylight saving time errors.
If errors are identified for either the time zone (TZ) or daylight saving
time (DST) settings, correct the time zone settings as per instructions in
Set the time zone on page 6-34
b.
Identify CARESCAPE Network IX and MC addressing scheme errors:
Verify the CARESCAPE Network IX and MC addressing schemes
match for all the CIC Pro centers on the CARESCAPE Network.
If the addressing schemes do not match, complete Network IP address
configuration for each CIC Pro center that requires IP address changes,
as per instructions in the appropriate CIC Pro Clinical Information
Center Service Manual.
4.
If there is any CIC Pro center hardware running v4.1.1 on the network, make sure
that the v4.1.1-1 patch is applied.
5.
Make sure that
only
the following compatible CIC Pro center versions co-exist in
a given
care area
:
CIC v5.1.x
CIC v5.0.x
CIC v4.1.1-1
CIC v4.0.7 or CIC v4.0.8
NOTE
CIC Pro center hardware running v1.5, or any version of Central Scope, can
exist independently in a separate
care area
, but not in the same care unit
name with the software versions listed above.
6.
If there is one or more Aware Gateway on the network, configure the Time
Master settings at the Aware Gateway. See the latest Aware Gateway Service
Manual for instructions. Consult Tech Support and hospital IT if any assistance is
required.
7.
Investigate if there is any unauthorized, non-GE medical equipment connected to
the MC/IX network. If any unauthorized non-GE equipment is found or reported
to be connected to the CARESCAPE Network, read and understand the
following Warning and communicate this information to the hospital IT/
biomedical or clinical staff before proceeding further:
Summary of Contents for CIC Pro
Page 12: ...x CIC Pro 2026419 033E ...
Page 13: ...2026419 033E CIC Pro 1 1 1 Introduction ...
Page 27: ...2026419 033E CIC Pro 2 1 2 Equipment overview ...
Page 53: ...2026419 033E CIC Pro 3 1 3 Licensing ...
Page 60: ...3 8 CIC Pro 2026419 033E Licensing ...
Page 61: ...2026419 033E CIC Pro 4 1 4 Service interfaces ...
Page 74: ...4 14 CIC Pro 2026419 033E Service interfaces ...
Page 75: ...2026419 033E CIC Pro 5 1 5 Installation ...
Page 94: ...5 20 CIC Pro 2026419 033E Installation ...
Page 95: ...2026419 033E CIC Pro 6 1 6 Configuration ...
Page 187: ...2026419 033E CIC Pro 7 1 7 Checkout procedures ...
Page 208: ...7 22 CIC Pro 2026419 033E Checkout procedures ...
Page 209: ...2026419 033E CIC Pro 8 1 8 Troubleshooting ...
Page 241: ...2026419 033E CIC Pro 9 1 9 Field replaceable units FRUs ...
Page 243: ...Field replaceable units FRUs 2026419 033E CIC Pro 9 3 Hard drive Flash drive ...
Page 244: ...9 4 CIC Pro 2026419 033E Field replaceable units FRUs Rack mount ...
Page 245: ...Field replaceable units FRUs 2026419 033E CIC Pro 9 5 Mounting options ...
Page 265: ...2026419 033E CIC Pro 10 1 10 Preventive maintenance ...
Page 279: ...2026419 033E CIC Pro 11 1 11 Reload software ...
Page 288: ...11 10 CIC Pro 2026419 033E Reload software ...
Page 289: ...2026419 033E CIC Pro 12 1 12 Upgrade software ...
Page 298: ...12 10 CIC Pro 2026419 033E Upgrade software ...
Page 299: ...2026419 033E CIC Pro A 1 A Electromagnetic compatibility ...
Page 305: ...2026419 033E CIC Pro B 1 B Backup and restore backed up data modules ...
Page 309: ...Device compatibility 2026419 033E CIC Pro C 1 C Device compatibility ...
Page 312: ...C 4 CIC Pro 2026419 033E Device compatibility ...
Page 313: ......