9-14
CIC Pro™
2026419-033E
Field replaceable units (FRUs)
3.
Lift the cover off the chassis.
Replacing the CPU PCB (mother board)
1.
Prepare the unit for disassembly on page 9-12
.
2.
.
3.
Remove the speaker cable from the CPU PCB.
4.
Remove the two SATA cables from the CPU PCB.
5.
Remove the LED light pipe from the CPU PCB.
6.
Remove the power supply cable from the CPU PCB.
7.
Remove the two fan cables from the CPU PCB.
8.
Remove the two serial cables from the CPU PCB.
9.
Remove the four screws that connect the DVI ports to the chassis (retain the
screws).
10. Remove the six screws that connect the CPU PCB to the chassis (retain the
screws).
11. Replace the CPU PCB.
NOTE
Do not discard the memory cards (qty 2) and battery.
12. Reverse steps to re-assemble the CPU.
13. Replace the original memory cards (qty 2) and battery.
14.
Re-assemble the cover on page 9-24
15. Perform the following checkout procedures:
Perform
Ground continuity test on page 10-6
Perform
Ground (earth) wire leakage current test on page 10-7
.
Perform
Enclosure leakage current test on page 10-8
.
Perform time adjustment.
Summary of Contents for CIC Pro
Page 12: ...x CIC Pro 2026419 033E ...
Page 13: ...2026419 033E CIC Pro 1 1 1 Introduction ...
Page 27: ...2026419 033E CIC Pro 2 1 2 Equipment overview ...
Page 53: ...2026419 033E CIC Pro 3 1 3 Licensing ...
Page 60: ...3 8 CIC Pro 2026419 033E Licensing ...
Page 61: ...2026419 033E CIC Pro 4 1 4 Service interfaces ...
Page 74: ...4 14 CIC Pro 2026419 033E Service interfaces ...
Page 75: ...2026419 033E CIC Pro 5 1 5 Installation ...
Page 94: ...5 20 CIC Pro 2026419 033E Installation ...
Page 95: ...2026419 033E CIC Pro 6 1 6 Configuration ...
Page 187: ...2026419 033E CIC Pro 7 1 7 Checkout procedures ...
Page 208: ...7 22 CIC Pro 2026419 033E Checkout procedures ...
Page 209: ...2026419 033E CIC Pro 8 1 8 Troubleshooting ...
Page 241: ...2026419 033E CIC Pro 9 1 9 Field replaceable units FRUs ...
Page 243: ...Field replaceable units FRUs 2026419 033E CIC Pro 9 3 Hard drive Flash drive ...
Page 244: ...9 4 CIC Pro 2026419 033E Field replaceable units FRUs Rack mount ...
Page 245: ...Field replaceable units FRUs 2026419 033E CIC Pro 9 5 Mounting options ...
Page 265: ...2026419 033E CIC Pro 10 1 10 Preventive maintenance ...
Page 279: ...2026419 033E CIC Pro 11 1 11 Reload software ...
Page 288: ...11 10 CIC Pro 2026419 033E Reload software ...
Page 289: ...2026419 033E CIC Pro 12 1 12 Upgrade software ...
Page 298: ...12 10 CIC Pro 2026419 033E Upgrade software ...
Page 299: ...2026419 033E CIC Pro A 1 A Electromagnetic compatibility ...
Page 305: ...2026419 033E CIC Pro B 1 B Backup and restore backed up data modules ...
Page 309: ...Device compatibility 2026419 033E CIC Pro C 1 C Device compatibility ...
Page 312: ...C 4 CIC Pro 2026419 033E Device compatibility ...
Page 313: ......