Configuration
2026419-033E
CIC Pro™
6-61
7.
Enter the
Automatic configuration
script or
Proxy server
address and port
according to the hospital IT Administrator. Click
OK
.
8.
Under
Automatic Configuration
, uncheck
Automatically detect settings
and
check
Use automatic configuration script options
.
9.
Click
OK
to close the
Local Area Network (LAN) settings
window.
10. Click
OK
again to close the
Internet Properties
window.
11. Confirm proper operation of the browser and the CARESCAPE Network by
completing the respective checkout procedures,
browser functionality on page 7-7
Check access to all other care units on page 7-
Check full disclosure report printing on page 7-9
.
Configure customize groupings
You can configure customized groups of parameters for viewing patient data.
NOTE
You can delete the default groups.
You can configure up to 12 custom groupings for patient data.
Set up customize groupings for graphic trends
1.
Log on to Setup CIC with service access on page 4-5
.
2.
From the multi-patient viewer, select any patient window.
3.
Click
Patient Data
.
4.
Click
Graphic Trends
.
Summary of Contents for CIC Pro
Page 12: ...x CIC Pro 2026419 033E ...
Page 13: ...2026419 033E CIC Pro 1 1 1 Introduction ...
Page 27: ...2026419 033E CIC Pro 2 1 2 Equipment overview ...
Page 53: ...2026419 033E CIC Pro 3 1 3 Licensing ...
Page 60: ...3 8 CIC Pro 2026419 033E Licensing ...
Page 61: ...2026419 033E CIC Pro 4 1 4 Service interfaces ...
Page 74: ...4 14 CIC Pro 2026419 033E Service interfaces ...
Page 75: ...2026419 033E CIC Pro 5 1 5 Installation ...
Page 94: ...5 20 CIC Pro 2026419 033E Installation ...
Page 95: ...2026419 033E CIC Pro 6 1 6 Configuration ...
Page 187: ...2026419 033E CIC Pro 7 1 7 Checkout procedures ...
Page 208: ...7 22 CIC Pro 2026419 033E Checkout procedures ...
Page 209: ...2026419 033E CIC Pro 8 1 8 Troubleshooting ...
Page 241: ...2026419 033E CIC Pro 9 1 9 Field replaceable units FRUs ...
Page 243: ...Field replaceable units FRUs 2026419 033E CIC Pro 9 3 Hard drive Flash drive ...
Page 244: ...9 4 CIC Pro 2026419 033E Field replaceable units FRUs Rack mount ...
Page 245: ...Field replaceable units FRUs 2026419 033E CIC Pro 9 5 Mounting options ...
Page 265: ...2026419 033E CIC Pro 10 1 10 Preventive maintenance ...
Page 279: ...2026419 033E CIC Pro 11 1 11 Reload software ...
Page 288: ...11 10 CIC Pro 2026419 033E Reload software ...
Page 289: ...2026419 033E CIC Pro 12 1 12 Upgrade software ...
Page 298: ...12 10 CIC Pro 2026419 033E Upgrade software ...
Page 299: ...2026419 033E CIC Pro A 1 A Electromagnetic compatibility ...
Page 305: ...2026419 033E CIC Pro B 1 B Backup and restore backed up data modules ...
Page 309: ...Device compatibility 2026419 033E CIC Pro C 1 C Device compatibility ...
Page 312: ...C 4 CIC Pro 2026419 033E Device compatibility ...
Page 313: ......