8-6
CIC Pro™
2026419-033E
Troubleshooting
Alarms are silenced for out-of-unit beds at
the CIC Pro center
Even though the setflags option
all_mvaudio
is set to
on
at the central
station, alarms may have been silenced.
Since this capability exists on the CIC Pro
center, care must be exercised not to
silence the alarm when setflags
-all_mvaudio
is configured to be
on
.
CIC Pro center continuously reboots during
the USB reload procedure
The reload process started without
disconnecting the network cables from the
CIC Pro center.
Disconnect all the network cables before
starting the reload process.
Loss of browser functionality
Parsing defect with Microsoft Internet
Explorer
Level of security prevents certain
applications from running
See Browser connectivity issues on page 8-
9.
Printer button dimmed and unselectable
when viewing stored patient data
Laser printer is not configured correctly
Restricted support from the Bedside
monitor
See Laser printer on page 8-9.
Unable to print to the Digital Writer
Digital writer not powered up
Printer door is open
Printer is out of paper
Printer is not configured correctly
See Digital display writer on page 8-10.
Unable to use the MultiKM feature
MultiKM license is missing
MultiKM feature is not configured
Activate the MultiKM license and perform
the required configuration as described in
Perform MultiKM (Multimouse) setup on
page 6-77
MultiKM worked for only one hour
MultiKM enable has failed
See MultiKM issues on page 8-10.
When using Multimouse, unable to locate
the control/focus screen
Cannot physically locate due to the physical
layout of the displays
Use
Ctrl+F1
to change the focus to the CIC
Pro center where the keyboard and mouse
are connected.
Unable to access CIC Pro center within a
Multimouse group
Incompatible software versions within the
Multimouse group
See Perform MultiKM (Multimouse) setup
on page 6-9.
Blank display screen
Abrupt Power failure
No UPS
Loose cables/connectors/connections
Internal component is loose
Blue display screen
Windows Operating system has
encountered a functional error
Red display screen
CIC Pro Application is starting up
Incorrect display colors
Issue with the video
See Check video function and status of
video card and drivers on page 7-19.
Unable to communicate with a device
See Unable to communicate with a device
on page 8-11.
Need to stop full disclosure collection at a
CIC Pro center running v5.1.x or later
See Set full disclosure mode on page 8-13.
Need to Ping full disclosure server
See Ping full disclosure server on page 8-
14.
Need to get information regarding full
disclosure data collection CIC Pro center
units
See List full disclosure beds on page 8-12.
Message/Symptom/Issue
Probable cause
Recommended action
Summary of Contents for CIC Pro
Page 12: ...x CIC Pro 2026419 033E ...
Page 13: ...2026419 033E CIC Pro 1 1 1 Introduction ...
Page 27: ...2026419 033E CIC Pro 2 1 2 Equipment overview ...
Page 53: ...2026419 033E CIC Pro 3 1 3 Licensing ...
Page 60: ...3 8 CIC Pro 2026419 033E Licensing ...
Page 61: ...2026419 033E CIC Pro 4 1 4 Service interfaces ...
Page 74: ...4 14 CIC Pro 2026419 033E Service interfaces ...
Page 75: ...2026419 033E CIC Pro 5 1 5 Installation ...
Page 94: ...5 20 CIC Pro 2026419 033E Installation ...
Page 95: ...2026419 033E CIC Pro 6 1 6 Configuration ...
Page 187: ...2026419 033E CIC Pro 7 1 7 Checkout procedures ...
Page 208: ...7 22 CIC Pro 2026419 033E Checkout procedures ...
Page 209: ...2026419 033E CIC Pro 8 1 8 Troubleshooting ...
Page 241: ...2026419 033E CIC Pro 9 1 9 Field replaceable units FRUs ...
Page 243: ...Field replaceable units FRUs 2026419 033E CIC Pro 9 3 Hard drive Flash drive ...
Page 244: ...9 4 CIC Pro 2026419 033E Field replaceable units FRUs Rack mount ...
Page 245: ...Field replaceable units FRUs 2026419 033E CIC Pro 9 5 Mounting options ...
Page 265: ...2026419 033E CIC Pro 10 1 10 Preventive maintenance ...
Page 279: ...2026419 033E CIC Pro 11 1 11 Reload software ...
Page 288: ...11 10 CIC Pro 2026419 033E Reload software ...
Page 289: ...2026419 033E CIC Pro 12 1 12 Upgrade software ...
Page 298: ...12 10 CIC Pro 2026419 033E Upgrade software ...
Page 299: ...2026419 033E CIC Pro A 1 A Electromagnetic compatibility ...
Page 305: ...2026419 033E CIC Pro B 1 B Backup and restore backed up data modules ...
Page 309: ...Device compatibility 2026419 033E CIC Pro C 1 C Device compatibility ...
Page 312: ...C 4 CIC Pro 2026419 033E Device compatibility ...
Page 313: ......