Combo mode Example 3:
1.
A telemetry bed BED* is admitted and full disclosure data is collected.
2.
A hardwire bed BED is admitted and full disclosure data is collected. Two
separate full disclosure data stores exist, one for each bed (possibly on a different
CIC Pro centers).
3.
The two beds are put into Combo mode. When this happens, BED* data is kept
because it is older than BED data, and BED data is deleted. Full disclosure data
collection continues, and the data is stored under the name BED.
4.
Combo mode is broken by discharging BED*. Full disclosure data collection
continues, and the data is still stored under the name BED.
Combo mode Example 4:
1.
A telemetry bed BED* is admitted and full disclosure data is collected.
2.
A hardwired bed BED is admitted and full disclosure data is collected. Two
separate full disclosure data stores exist, one for each bed (possibly on a different
CIC Pro center).
3.
The two beds are put into Combo mode. When this happens, BED* data is kept
because it is older than BED data, and BED data is deleted. Full disclosure data
collection continues, and the data is stored under the name BED.
4.
Combo mode is broken by discharging BED. Full disclosure data collection
continues, and the data is now stored under the name BED*.
Bed name/IP address changes (rover rule)
The CIC Pro center tracks full disclosure data for individual beds using a combination
of the bed name and IP address.
If a bed name changes during full disclosure data collection, information within the
data reflects the new name at the point the name change occurred.
If a bed IP address changes during full disclosure data collection, but the name does
not, it is interpreted as a new and different bed, and a new full disclosure data store is
created and utilized for the new bed.
Summary of Contents for CIC Pro
Page 12: ...x CIC Pro 2026419 033E ...
Page 13: ...2026419 033E CIC Pro 1 1 1 Introduction ...
Page 27: ...2026419 033E CIC Pro 2 1 2 Equipment overview ...
Page 53: ...2026419 033E CIC Pro 3 1 3 Licensing ...
Page 60: ...3 8 CIC Pro 2026419 033E Licensing ...
Page 61: ...2026419 033E CIC Pro 4 1 4 Service interfaces ...
Page 74: ...4 14 CIC Pro 2026419 033E Service interfaces ...
Page 75: ...2026419 033E CIC Pro 5 1 5 Installation ...
Page 94: ...5 20 CIC Pro 2026419 033E Installation ...
Page 95: ...2026419 033E CIC Pro 6 1 6 Configuration ...
Page 187: ...2026419 033E CIC Pro 7 1 7 Checkout procedures ...
Page 208: ...7 22 CIC Pro 2026419 033E Checkout procedures ...
Page 209: ...2026419 033E CIC Pro 8 1 8 Troubleshooting ...
Page 241: ...2026419 033E CIC Pro 9 1 9 Field replaceable units FRUs ...
Page 243: ...Field replaceable units FRUs 2026419 033E CIC Pro 9 3 Hard drive Flash drive ...
Page 244: ...9 4 CIC Pro 2026419 033E Field replaceable units FRUs Rack mount ...
Page 245: ...Field replaceable units FRUs 2026419 033E CIC Pro 9 5 Mounting options ...
Page 265: ...2026419 033E CIC Pro 10 1 10 Preventive maintenance ...
Page 279: ...2026419 033E CIC Pro 11 1 11 Reload software ...
Page 288: ...11 10 CIC Pro 2026419 033E Reload software ...
Page 289: ...2026419 033E CIC Pro 12 1 12 Upgrade software ...
Page 298: ...12 10 CIC Pro 2026419 033E Upgrade software ...
Page 299: ...2026419 033E CIC Pro A 1 A Electromagnetic compatibility ...
Page 305: ...2026419 033E CIC Pro B 1 B Backup and restore backed up data modules ...
Page 309: ...Device compatibility 2026419 033E CIC Pro C 1 C Device compatibility ...
Page 312: ...C 4 CIC Pro 2026419 033E Device compatibility ...
Page 313: ......