Preventive maintenance
2026419-033E
CIC Pro™
10-7
2.
Connect the negative (-) lead of the DMM to the protective earth terminal
(ground pin in power inlet connector) or the protective earth pin in the Mains
plug (ground pin in power cord).
3.
Set the DMM to the milliohm (mW) range.
4.
Connect the positive (+) lead of the DMM to the equipotential plug on the CIC
Pro center.
5.
Resistance must read:
0.1 ohm or less without power cord
0.2 ohms or less with power cord
Impedance of protective earth connection test
This test, unlike a ground continuity test, will also stress the ground system by using
special ground bond testers. This test normally is only required as a manufacturing
production test to receive safety agency compliance. Some country agencies do
require this test after field equipment repairs (i.e., Germany's DIN VDE 0751
standards). Consult your country/local safety agency if in question.
1.
A current of 25A from a current source with a frequency of 50 or 60 Hz with a
no-load voltage not exceeding 6 V is passed for at least 5 s through the protective
earth terminal or the protective earth pin in the mains plug and each accessible
metal part which could become live in case of failure in basic insulation.
2.
The voltage drop between the parts described is measured and the impedance
determined from the current and voltage drop. It shall not exceed the values
indicated.
For equipment without a power supply cord the impedance between the protective
earth terminal and any accessible metal part which is protectively earthed shall not
exceed 0.1 ohms.
For equipment with a power supply cord, the impedance between the protective earth
pin in the mains plug and any accessible metal part which is protectively earthed shall
not exceed 0.2 ohms.
When taking this measurement, move the unit's power cord around. There should be
no fluctuations in resistance.
Ground (earth) wire leakage current test
NOTE
The DMM plus leakage tester network shown is the circuitry defined by the IEC/
EN/UL 60601-1 standard for measuring leakage current.
Perform this test to measure current leakage through the ground (earth) wire of the
equipment during normal operation.
Summary of Contents for CIC Pro
Page 12: ...x CIC Pro 2026419 033E ...
Page 13: ...2026419 033E CIC Pro 1 1 1 Introduction ...
Page 27: ...2026419 033E CIC Pro 2 1 2 Equipment overview ...
Page 53: ...2026419 033E CIC Pro 3 1 3 Licensing ...
Page 60: ...3 8 CIC Pro 2026419 033E Licensing ...
Page 61: ...2026419 033E CIC Pro 4 1 4 Service interfaces ...
Page 74: ...4 14 CIC Pro 2026419 033E Service interfaces ...
Page 75: ...2026419 033E CIC Pro 5 1 5 Installation ...
Page 94: ...5 20 CIC Pro 2026419 033E Installation ...
Page 95: ...2026419 033E CIC Pro 6 1 6 Configuration ...
Page 187: ...2026419 033E CIC Pro 7 1 7 Checkout procedures ...
Page 208: ...7 22 CIC Pro 2026419 033E Checkout procedures ...
Page 209: ...2026419 033E CIC Pro 8 1 8 Troubleshooting ...
Page 241: ...2026419 033E CIC Pro 9 1 9 Field replaceable units FRUs ...
Page 243: ...Field replaceable units FRUs 2026419 033E CIC Pro 9 3 Hard drive Flash drive ...
Page 244: ...9 4 CIC Pro 2026419 033E Field replaceable units FRUs Rack mount ...
Page 245: ...Field replaceable units FRUs 2026419 033E CIC Pro 9 5 Mounting options ...
Page 265: ...2026419 033E CIC Pro 10 1 10 Preventive maintenance ...
Page 279: ...2026419 033E CIC Pro 11 1 11 Reload software ...
Page 288: ...11 10 CIC Pro 2026419 033E Reload software ...
Page 289: ...2026419 033E CIC Pro 12 1 12 Upgrade software ...
Page 298: ...12 10 CIC Pro 2026419 033E Upgrade software ...
Page 299: ...2026419 033E CIC Pro A 1 A Electromagnetic compatibility ...
Page 305: ...2026419 033E CIC Pro B 1 B Backup and restore backed up data modules ...
Page 309: ...Device compatibility 2026419 033E CIC Pro C 1 C Device compatibility ...
Page 312: ...C 4 CIC Pro 2026419 033E Device compatibility ...
Page 313: ......