TROUBLESHOOTING: Data Acquisition Tests
Revision A
DASH 2000 Patient Monitor
4-11
2000 412-001
Generate Static BP
Waveform
Set the patient simulator BP output to 200 mmHg, static pressure.
¥
Verify the BP channel is working correctly if systolic, diastolic, and
mean pressure values for ART are displayed between 194 and
206 mmHg.
BP Waveform Does Not
Appear Correctly On The
Display
1.
If the BP waveform displayed on the monitor appear noisy or
distorted, test the Patient simulator and simulator test cables and on
a working monitor to determine the source of the problem.
2.
If the static pressure test results were inaccurate, test the Patient
simulator and simulator test cables and on a working monitor to
determine the source of the problem.
3.
If the patient simulator and associated test cables are determined to
be functioning correctly, the acquisition PCB is suspect. Swap the
acquisition PCB into a working monitor to determine if replacement
is necessary.
BP Waveform Does Not
Appear On The Display At
All
1.
If the ART parameter label, reading and associated waveform do not
display on the monitor, verify the patient simulator and associated
test cables on a working monitor.
2.
Inspect the BP front panel connector on the monitor for bent or
broken pins.
3.
Perform continuity tests between the front panel connector of the
monitor, front panel circuit assembly located behind the front panel
connector and connection to the acquisition PCB.
4.
If the patient simulator and associated test cables are determined to
be functioning correctly and the continuity tests yield no
malfunction, the acquisition PCB is suspect. Swap the acquisition
PCB into a working monitor to determine if replacement is necessary.
Respiration
Functions
1.
Connect the MEI Multifunction Microsimulator, pn. MARQ1, and
appropriate patient cables to the ECG/RESP front panel connector
on the monitor.
2.
Adjust the patient simulator to output a respiration waveform using
the following settings:
¥
Rate BPM Ð 30
¥
Baseline Impedance Ohms Ð 750,
¥
ÆR Ohms Ð 2.0.
3.
Enable the respiration function of the monitor:
¥
Select MONITOR SETUP from the main menu display on the
monitor,
¥
Select PARAMETERS ON/OFF from the monitor setup menu.
Next, turn and push the Trim Knob to:
¥
scroll to and select RR in the parameters on/off pop-up window.
¥
toggle and select ON in the RR line of the parameters on/off pop-
up window.
Summary of Contents for marquette Dash 2000
Page 12: ...Contents viii DASH 2000 Patient Monitor Revision A 2000 412 001 ...
Page 82: ...Maintenance Repair Log 3 36 DASH 2000 Patient Monitor Revision A 2000 412 001 ...
Page 156: ...UPPER LEVEL ASSEMBLY Spare Parts List 7 10 DASH 2000 Patient Monitor Revision A 2000 412 001 ...
Page 157: ...Revision A DASH 2000 Patient Monitor 8 1 2000 412 001 8 ASSEMBLY DRAWINGS Contents ...
Page 159: ......
Page 173: ......
Page 178: ......
Page 180: ......
Page 181: ......
Page 182: ......
Page 184: ......
Page 196: ......
Page 197: ......
Page 198: ......
Page 200: ......
Page 201: ......