UPPER LEVEL ASSEMBLY: Disassembly Procedure
7-4
DASH 2000 Patient Monitor
Revision A
2000 412-001
DASH 2000
Assembly
1.
Open the unit as described in
Opening Unit
.
2.
Remove the 2-pole converter plug and the flat cable connector for the
display from the PCB Mainboard.
3.
Tilt the display frame with the DAS board towards the front.
4.
Disconnect the DAS flat cable connector.
5.
Loosen the DAS input assembly.
6.
Remove NIBP tubing from DAS input.
7.
Remove Alarm Light cable from PCB Main Connector
For devices equipped with writer:
8.
Open writer and remove paper.
9.
Loosen the 2 screws in the writer housing and carefully pull out
the writer.
10. Remove writer cable which is attached to the PCB Mainboard by
means of a screw.
11. Loosen writer bracket and carefully pull it out towards the
inside.
12. Remove speaker from PCB Mainboard.
13. Remove 4 screws at the rear panel.
14. Remove 4 screws at the foot pad.
15. Remove 2 screws at the bottom of the enclosure.
16. Remove ground wire from enclosure.
17. Carefully pull the DASH2000 assembly from the enclosure
PCB Main Connector
1.
Disassemble the
DASH 2000 Assembly
as described above.
2.
Undo the 3 screws from PCB Main Connector and remove it.
PCB NIBP
1.
Disassemble the
DASH 2000 Assembly
as described above.
2.
Disconnect battery connector and remove the 2 screws from the
battery bracket.
3.
Disconnect the NIBP connector from the Mainboard.
4.
Undo the 3 screws which hold the NIBP board on the battery
bracket.
5.
Remove NIBP tubing from the pump.
6.
Disconnect NIBP Pump connector and remove NIBP board.
PCB Mainboard
1.
Disassemble the
DASH 2000 Assembly
as described above.
2.
Disconnect battery connector and remove the 2 screws from the
battery bracket.
3.
Disconnect the NIBP connector from the Mainboard and remove the
battery bracket with NIBP board.
4.
Remove the battery and the battery pad.
Summary of Contents for marquette Dash 2000
Page 12: ...Contents viii DASH 2000 Patient Monitor Revision A 2000 412 001 ...
Page 82: ...Maintenance Repair Log 3 36 DASH 2000 Patient Monitor Revision A 2000 412 001 ...
Page 156: ...UPPER LEVEL ASSEMBLY Spare Parts List 7 10 DASH 2000 Patient Monitor Revision A 2000 412 001 ...
Page 157: ...Revision A DASH 2000 Patient Monitor 8 1 2000 412 001 8 ASSEMBLY DRAWINGS Contents ...
Page 159: ......
Page 173: ......
Page 178: ......
Page 180: ......
Page 181: ......
Page 182: ......
Page 184: ......
Page 196: ......
Page 197: ......
Page 198: ......
Page 200: ......
Page 201: ......