CALIBRATION: Noninvasive Blood Pressure
Revision A
DASH 2000 Patient Monitor
5-5
2000 412-001
Calibration
Procedure
1.
Remove all cables except for the power cord from the monitor.
2.
Apply power to the monitor.
¥
Plug the power cord into a working ac power wall receptacle and
turn the monitor on.
3.
Use the Trim Knob control to scroll to MONITOR SETUP in the
monitor main menu and press the Trim Knob control to select it.
4.
Use the Trim Knob control to scroll to SERVICE MODE in the
monitor setup menu and press the Trim Knob control to select it.
5.
A service menu password window appears on the monitor display, as
shown in the figure at the left. A password is required to prevent
non-service personnel from accessing the service menus. The
password is four numbers that represent the date that currently
resides in a memory circuit within the monitor (please note that this
may or may not be the correct date). In the password, the first two
numbers, starting from the left, represent the day and the second
two numbers represent the month of whatever date that currently
resides in the memory circuits of the monitor. For example, the
seventh day of the third month (March 7th) would be represented in
the password as 0703 (
ddmm
). Note the date that is currently on the
monitor display and follow these steps to enter the password:
◆
Rotate the Trim Knob control to highlight the password number
that you would like to change.
◆
To change the highlighted number, press the Trim Knob control.
◆
Rotate the Trim Knob control until the correct number is
displayed in the selected field.
◆
To enter the number, press the Trim Knob control.
◆
Repeat these steps until all password numbers are correctly
displayed.
◆
Once you have entered the correct password numbers, rotate the
Trim Knob control to highlight SERVICE MODE in the enter
password window.
◆
Press the Trim Knob control one more time to enter the password
and access the service menus of the monitor.
MAIN
MENU
ALARM
CONTROL
PATIENT
DATA
ADMIT
MENU
MONITOR
SETUP
MAIN
MENU
MONITOR
DEFAULTS
DISPLAY
SETUP
GRAPH
SETUP
PREV.
MENU
PARAM.
ON / OFF
SOFTWARE
SUMMARY
SERVICE
MODE
SERVICE
MODE
6
0
1
0
Summary of Contents for marquette Dash 2000
Page 12: ...Contents viii DASH 2000 Patient Monitor Revision A 2000 412 001 ...
Page 82: ...Maintenance Repair Log 3 36 DASH 2000 Patient Monitor Revision A 2000 412 001 ...
Page 156: ...UPPER LEVEL ASSEMBLY Spare Parts List 7 10 DASH 2000 Patient Monitor Revision A 2000 412 001 ...
Page 157: ...Revision A DASH 2000 Patient Monitor 8 1 2000 412 001 8 ASSEMBLY DRAWINGS Contents ...
Page 159: ......
Page 173: ......
Page 178: ......
Page 180: ......
Page 181: ......
Page 182: ......
Page 184: ......
Page 196: ......
Page 197: ......
Page 198: ......
Page 200: ......
Page 201: ......