CALIBRATION: Noninvasive Blood Pressure
5-10
DASH 2000 Patient Monitor
Revision A
2000 412-001
as CUFF in the NBP parameter box) on the monitor and manometer
are equal (± 1 mmHg) for
at least
one full minute.
11. Rotate the Trim Knob control to highlight CHECK CAL IN
PROGRESS and press the Trim Knob control to select it.
12. Rotate the Trim Knob control to highlight STOP and press the Trim
Knob control to select it. The monitor automatically releases
pneumatic pressure in the entire plumbing circuit.
13. Turn the monitor off, turn the manometer off and remove the test
apparatus from the monitor.
ECG
150/ 50
21-JUN-1999 16:27
II
V5
I
ICU-BED 5
MARTIN
X
NBP
200/ 80
S
CUFF
X
250
X /
MAIN
MENU
PREV.
MENU
CAL
ZERO:
OFF
CHECK
CAL:
IN PROG
CAL
GAIN:
OFF
MAIN
MENU
CAL
ZERO:
OFF
CHECK
CAL:
IN PROG.
PREV.
MENU
CAL
GAIN:
OFF
>START
>
STOP
CHECK
CAL:
IN PROG.
Summary of Contents for marquette Dash 2000
Page 12: ...Contents viii DASH 2000 Patient Monitor Revision A 2000 412 001 ...
Page 82: ...Maintenance Repair Log 3 36 DASH 2000 Patient Monitor Revision A 2000 412 001 ...
Page 156: ...UPPER LEVEL ASSEMBLY Spare Parts List 7 10 DASH 2000 Patient Monitor Revision A 2000 412 001 ...
Page 157: ...Revision A DASH 2000 Patient Monitor 8 1 2000 412 001 8 ASSEMBLY DRAWINGS Contents ...
Page 159: ......
Page 173: ......
Page 178: ......
Page 180: ......
Page 181: ......
Page 182: ......
Page 184: ......
Page 196: ......
Page 197: ......
Page 198: ......
Page 200: ......
Page 201: ......