TROUBLESHOOTING: Battery Failure
Revision A
DASH 2000 Patient Monitor
4-5
2000 412-001
Opening the Device and
Battery Replacement
1.
Disassemble the
DASH 2000 Assembly
as described in
"Disassembly Procedure".
2.
Disconnect battery connector and remove the 2 screws from the
battery bracket.
3.
Remove old battery and insert the new battery.
4.
Assembly the device by reversing the above operating steps.
Due to replacement of the battery during which the battery management
system was without current the fuel gauge no longer reflects the current
battery status. The system indicates BATTERY LOW.
Everytime, after the battery was disconnected from the circuit
board drain the battery by operating the device on battery
power until it switches off. Then connect the device to the power
line and start a conditioning cycle.
NOTE
Disposal Notice:
Should this product (battery) become
damaged beyond repair, or for some reason its useful life
is considered to be at an end, please observe all local,
state, and federal regulation that relate to the disposal of
products that contain lead, batteries, plastics, etc.
Summary of Contents for marquette Dash 2000
Page 12: ...Contents viii DASH 2000 Patient Monitor Revision A 2000 412 001 ...
Page 82: ...Maintenance Repair Log 3 36 DASH 2000 Patient Monitor Revision A 2000 412 001 ...
Page 156: ...UPPER LEVEL ASSEMBLY Spare Parts List 7 10 DASH 2000 Patient Monitor Revision A 2000 412 001 ...
Page 157: ...Revision A DASH 2000 Patient Monitor 8 1 2000 412 001 8 ASSEMBLY DRAWINGS Contents ...
Page 159: ......
Page 173: ......
Page 178: ......
Page 180: ......
Page 181: ......
Page 182: ......
Page 184: ......
Page 196: ......
Page 197: ......
Page 198: ......
Page 200: ......
Page 201: ......