EQUIPMENT OVERVIEW: Performance Specifications
2-10
DASH 2000 Patient Monitor
Revision A
2000 412-001
Analog Output
DeÞbrillator
Synchronization Pulse
Environmental
SpeciÞcations
ECG:
Gain:
1 V/mV ±10%
DC offset:
±100 mV (max)
Noise:
<5 mVp-p (0-300 Hz)
Frequency response:
0.05 Hz to 100 Hz +7/-0 Hz
Marker out:
Time delay:
35 ms (max), R-wave peak to
leading edge of pulse.
Amplitude selectable in Service Menu
+5 V selection:
3.5 V (min) at 1 mA sourcing; 0.5 V
(max) at 5 mA sinking.
+12 V selection:
11.0 V (max) at 1 mA sourcing;
0.75 V (max) at 5 mA sinking.
Pulse width:
10 ms ±10% or 100 ms ±10% in
Service Menu
Output impedance:
50 Ohms nominal
Current limit:
15 mA nominal, both sourcing and
sinking.
Marker in:
Input threshold:
VIH = +2.5 V (min);
VIL = +1.5 V (max)
Input hysteresis:
650 mV typical
Maximum input voltage:
±30 V (with respect to ground on
pin 2)
Input impedance:
10 k (min) for Ð25 V < Vin < 25 V
Pulse width:
1.0 ms (min), Vin < 2.5 V
Power requirements:
AC voltage:
100 Ð 240 VAC ±10%
Power consumption:
16 W normal use, 45 W fast charge
Cooling:
Convection
Heat dissipation:
240 BTU/hr
Battery:
nickel-cadmium (NiCd), 12 V,
2.0 ampere hours
Fuses:
100 Ð 240 VAC: T2.0A, 250 VAC,
5 x 20 mm
Design (general):
Continuous, not protected against
ingress of liquids
Battery operation time:
General:
Battery age will affect operating
time. SpO2 and NBP monitoring, as
well as battery age, reduce
operating time.
Monochrome LCD display:
Typical operation time while
monitoring ECG is 3.5 hours from a
new, fully-charged battery.
Color LCD display:
Typical operation time while
monitoring ECG is 3 hours from a
new, fully-charged battery.
Min. battery operating time:
1.5 hours
Battery charge time to 90%:
1 hour to 3 hours
Summary of Contents for marquette Dash 2000
Page 12: ...Contents viii DASH 2000 Patient Monitor Revision A 2000 412 001 ...
Page 82: ...Maintenance Repair Log 3 36 DASH 2000 Patient Monitor Revision A 2000 412 001 ...
Page 156: ...UPPER LEVEL ASSEMBLY Spare Parts List 7 10 DASH 2000 Patient Monitor Revision A 2000 412 001 ...
Page 157: ...Revision A DASH 2000 Patient Monitor 8 1 2000 412 001 8 ASSEMBLY DRAWINGS Contents ...
Page 159: ......
Page 173: ......
Page 178: ......
Page 180: ......
Page 181: ......
Page 182: ......
Page 184: ......
Page 196: ......
Page 197: ......
Page 198: ......
Page 200: ......
Page 201: ......