Maintenance: Checkout Procedures
3-16
DASH 2000 Patient Monitor
Revision A
2000 412-001
Verify defib sync markers
3.
Attach a jumper wire between pin-1 (Marker Out) and pin-2 (Marker
In) of the DEFIB SYNC connector located on the front of the monitor.
Verify an X is displayed in each of the QRS Complex (ECG waveform)
R-Waves on the monitor display, similar to those shown in the
illustration below.
Defibrillator synchronization
tests completion
4.
Remove the jumper wire installed in the previous step from the
DEFIB SYNC connector. This completes the defibrillator
synchronization tests.
DEFIB SYNC connector:
Marker Out (frequency)
Signal Pin:
1
Ground Pin:
4
Time/Division:
0.2 s
Volts/Division:
1 V, when
Defi Sync output configured
for 5 V
DEFIB SYNC connector:
Marker Out (pulse width)
Signal Pin:
1
Ground Pin:
4
Time/Division:
5 ms
Volts/Division:
1 V, when
Defi Sync output configured
for 5 V and 10 ms
X
X
X
X
Observe normal
R-waves before
the jumper is
installed.
Observe the X in the
R-waves while the
jumper is installed.
Summary of Contents for marquette Dash 2000
Page 12: ...Contents viii DASH 2000 Patient Monitor Revision A 2000 412 001 ...
Page 82: ...Maintenance Repair Log 3 36 DASH 2000 Patient Monitor Revision A 2000 412 001 ...
Page 156: ...UPPER LEVEL ASSEMBLY Spare Parts List 7 10 DASH 2000 Patient Monitor Revision A 2000 412 001 ...
Page 157: ...Revision A DASH 2000 Patient Monitor 8 1 2000 412 001 8 ASSEMBLY DRAWINGS Contents ...
Page 159: ......
Page 173: ......
Page 178: ......
Page 180: ......
Page 181: ......
Page 182: ......
Page 184: ......
Page 196: ......
Page 197: ......
Page 198: ......
Page 200: ......
Page 201: ......