TROUBLESHOOTING: Service Mode Menu
Revision A
DASH 2000 Patient Monitor
4-19
2000 412-001
Severity of the Error
Severity is a measure of how the event/error affected the system. There
are three levels of severity. The following is a list of these levels
accompanied by a brief description of each:
Continue: Ñ
The event or error was logged, the task may or may not
have completed, but the system was able to continue operating. Most
error log entries have this severity level.
Fatal: Ñ
The event or error was logged, the task did not complete, and
the system was unable to continue operating as recovery was not
possible. This level of severity in an event or error is always followed
by an automatic warm start.
Forced Restart: Ñ
The operating system restarted normally after a
known condition, such as an Internet address change, patient
discharge, etc.