Maintenance: Checkout Procedures
Revision A
DASH 2000 Patient Monitor
3-15
2000 412-001
DeÞbrillator
Synchronization
Tests
1.
Use the figure at the left as a reference for connecting the
oscilloscope to the DEFIB SYNC connector, located on the back panel
of the monitor, for performing these tests.
2.
Test the ECG and Marker Out signals from the DEFIB SYNC
connector. They should closely resemble the waveforms in the figures
below. Note that there are two Marker Out traces shown below. The
upper Marker Out figure references the frequency aspects of the
signal. The lower Marker Out figure references the pulse width
aspects of the signal.
Patient Simulator Setup:
HR Ð 80 bpm
HR amplitude Ð 1,0 mV
Display Setup:
Channel I Ð Lead II
PIN
Signal Name
I/O
Signal Description
1
MARKER_OUT
O
Digital defibrillator output
synchronization signal
2
MARKER_IN
I
Digital defibrillator input signal
3
ANALOG_GND
Ð
Analog return
4
DIGITAL_GND
Ð
Digital return
5
NC
6
NC
7
ECG_OUT
O
Analog ECG output signal
DEFIB SYNC connector:
ECG
Signal Pin:
7
Ground Pin:
3
Time/Division:
0.2 s
Volts/Division:
0.5 V
Summary of Contents for marquette Dash 2000
Page 12: ...Contents viii DASH 2000 Patient Monitor Revision A 2000 412 001 ...
Page 82: ...Maintenance Repair Log 3 36 DASH 2000 Patient Monitor Revision A 2000 412 001 ...
Page 156: ...UPPER LEVEL ASSEMBLY Spare Parts List 7 10 DASH 2000 Patient Monitor Revision A 2000 412 001 ...
Page 157: ...Revision A DASH 2000 Patient Monitor 8 1 2000 412 001 8 ASSEMBLY DRAWINGS Contents ...
Page 159: ......
Page 173: ......
Page 178: ......
Page 180: ......
Page 181: ......
Page 182: ......
Page 184: ......
Page 196: ......
Page 197: ......
Page 198: ......
Page 200: ......
Page 201: ......