EQUIPMENT OVERVIEW: Performance Specifications
Revision A
DASH 2000 Patient Monitor
2-9
2000 412-001
Pulse Oximetry
(SpO2)
Non-invasive Blood
Pressure (NBP)
Alarms:
User-selectable upper and lower
limits for systolic, diastolic, and
mean pressures
Parameters monitored:
Arterial oxygen saturation (SpO2)
and peripheral pulse rate (PPR)
SpO2 range:
calibrated:
50 Ð 100%
total:
0 Ð 100%
PPR range:
25 Ð 250 beats per minute (±3 beats
per minute)
Accuracy:
Actual accuracy depends on probe.
Please reference manufacturer's
specifications.
SpO2:
±2% (70 Ð 100% SpO2) ±1 standard
deviation
±3% (50 Ð 69% SpO2) ±1 standard
deviation
PPR
±3 beats per minute
Alarms:
User-selectable upper and lower
limits for SpO2 and PPR
Measurement technique:
Oscillometric
Displayed parameters:
Systolic, diastolic, and mean
pressures, pulse rate, time of last
measurement
Measurement modes:
Manual, auto, and stat
Heart rate detection:
30 to 300 beats per minute
Total cycle time:
20 to 40 seconds typical (dependent
on heart rate and motion artifact)
Automatic cycle times:
0 to 24 hours
Auto zero:
Zero pressure reference prior to
each cuff inflation
Tubing length:
Adult:
12 feet (3.6 m)
Neonatal:
8 feet (2.4 m)
Automatic cuff deflation:
Cycle time exceeding 3 minutes
(90 seconds neonatal), power off, or
cuff pressure exceeds 300 mmHg
(+10%) adult, 150 mmHg (+10%)
neonatal
Cuff sizes:
Disposable:
Large adult, adult, small adult,
pediatric, small pediatric, and
infant
Reusable:
Thigh, large adult, adult, child, and
infant
Alarms:
User-selectable upper and lower
limits for systolic, diastolic, and
mean pressures
Summary of Contents for marquette Dash 2000
Page 12: ...Contents viii DASH 2000 Patient Monitor Revision A 2000 412 001 ...
Page 82: ...Maintenance Repair Log 3 36 DASH 2000 Patient Monitor Revision A 2000 412 001 ...
Page 156: ...UPPER LEVEL ASSEMBLY Spare Parts List 7 10 DASH 2000 Patient Monitor Revision A 2000 412 001 ...
Page 157: ...Revision A DASH 2000 Patient Monitor 8 1 2000 412 001 8 ASSEMBLY DRAWINGS Contents ...
Page 159: ......
Page 173: ......
Page 178: ......
Page 180: ......
Page 181: ......
Page 182: ......
Page 184: ......
Page 196: ......
Page 197: ......
Page 198: ......
Page 200: ......
Page 201: ......