TROUBLESHOOTING: Service Tips
Revision A
DASH 2000 Patient Monitor
4-21
2000 412-001
Processor PCB
Symptoms
Symptoms with network communications, asynchronous
communications, NBP control, analog output, audio/sound generation,
and remote video signals/communications as well as other display-related
problems all may be associated with processor PCB failure. All of these are
functions controlled by microcontroller or graphics processing circuitry
located on the processor PCB.
Power Supply PCB
Symptoms
The power supply provides power that is used throughout the DASH
monitor. The supply voltages are generated for various applications on
the PCB Mainboard out of the +9 to +18 regulated VDC voltage. Below is
a list of the supply voltages and where and how these voltages are
applied. Problems in any of the following areas may be associated with
power supply failure.
+9 to +18 VDC
Supply Applications
¥
Voltage Converters (+12 VDC, -12 VDC, +5 VDC, +3.3 VDC,
+26 VDC, -23 VDC)
¥
Acquisition PCB
¥
Expansion Interface
¥
Writer Interface
¥
On/Off Interface
+12 VDC
Supply Applications
¥
Memory card (PCMCIA) slot Ð programming power source
¥
Defib marker out Ð power source for defib sync jack
¥
Audio amplifier Ð power source (speaker)
¥
NBP compressor (pump assembly) and solenoid valves Ð power source
+3.3 VDC
Supply Applications
¥
Main processor PCB Ð logic power source
+5 VDC
Supply Applications
¥
Main processor PCB Ð logic power source
¥
Display assembly Ð logic power source and backlight power source
¥
Ethernet transceiver
¥
Expansion Interface Ð logic power source
¥
Memory card (PCMCIA) slot Ð logic power source
¥
Remote display Ð logic power source
¥
Main memory Ð flash memory programming power source
+26 VDC
Supply Applications
¥
Color LCD voltage
-23 VDC
Supply Applications
¥
Monochrome LCD voltage
Summary of Contents for marquette Dash 2000
Page 12: ...Contents viii DASH 2000 Patient Monitor Revision A 2000 412 001 ...
Page 82: ...Maintenance Repair Log 3 36 DASH 2000 Patient Monitor Revision A 2000 412 001 ...
Page 156: ...UPPER LEVEL ASSEMBLY Spare Parts List 7 10 DASH 2000 Patient Monitor Revision A 2000 412 001 ...
Page 157: ...Revision A DASH 2000 Patient Monitor 8 1 2000 412 001 8 ASSEMBLY DRAWINGS Contents ...
Page 159: ......
Page 173: ......
Page 178: ......
Page 180: ......
Page 181: ......
Page 182: ......
Page 184: ......
Page 196: ......
Page 197: ......
Page 198: ......
Page 200: ......
Page 201: ......