CONFIGURATION: Load Software Over The Network
Revision A
DASH 2000 Patient Monitor
6-17
2000 412-001
Activate Software
You may download all the monitors in the care unit, then activate the
upgraded software.
After all of the files have been successfully loaded, press the monitor
power switch to OFF then ON to operate the monitor with the newly
loaded software.
Completion
Follow these steps to configure (setup) proper graph locations for an
internal or network writer:
1.
Use the Trim Knob to select the following menu options selection
sequence. Beginning at the monitor main menu options, scroll to and
select:
MONITOR SETUP
GRAPH SETUP
GRAPH LOCATION
2.
From the GRAPH LOCATION menu option items, use the Trim Knob
to scroll to and select:
a.
The MANUAL GRAPH LOCATION option item (the monitor
may take up to a minute to poll the network for available
writers), then choose one of the manual graph locations from that
list of writers; then
b.
The ALARM GRAPH LOCATION option item, then choose one of
the alarm graph locations from that list of writers; and finally
c.
The PRINT WINDOW LOCATION option item, then choose one
of the print window locations from that list of writers.
3.
Connect a patient simulator to the monitor. Admit and generate
patient waveforms at the monitor with the simulator powered up.
Perform the following steps to test the communication paths between
the monitor and each selected writer.
a.
Press the GRAPH GO/STOP key on the monitor front panel and
verify the graph output arrives at the selected manual graph
location. Press the GRAPH GO/STOP key again to stop the
manual graph.
b.
Cause a fatal alarm by switching the simulator power off and
verify the graph output arrives at the selected alarm graph
location.
c.
Bring up a non-real-time window on the monitor display and
print the window. Verify the print output arrives at the selected
print window location.
d. Verify software was downloaded successfully. Execute the
following menu option selection sequence, beginning at the
monitor main menu:
MONITOR SETUP
SOFTWARE SUMMARY
Press the GRAPH GO/STOP KEY
Compare displayed monitor software revisions with the those
previously printed or written down. Repeat the entire procedure if
software revisions are not properly updated.
Summary of Contents for marquette Dash 2000
Page 12: ...Contents viii DASH 2000 Patient Monitor Revision A 2000 412 001 ...
Page 82: ...Maintenance Repair Log 3 36 DASH 2000 Patient Monitor Revision A 2000 412 001 ...
Page 156: ...UPPER LEVEL ASSEMBLY Spare Parts List 7 10 DASH 2000 Patient Monitor Revision A 2000 412 001 ...
Page 157: ...Revision A DASH 2000 Patient Monitor 8 1 2000 412 001 8 ASSEMBLY DRAWINGS Contents ...
Page 159: ......
Page 173: ......
Page 178: ......
Page 180: ......
Page 181: ......
Page 182: ......
Page 184: ......
Page 196: ......
Page 197: ......
Page 198: ......
Page 200: ......
Page 201: ......