EQUIPMENT OVERVIEW: Preparation for Use
Revision A
DASH 2000 Patient Monitor
2-13
2000 412-001
Monitor Ventilation
Requirements
The monitor is capable of producing as much as 170 BTu per hour of heat
load. This is equivalent to approximately 50 watts of energy.
WARNING
Failure to properly ventilate the monitor may cause
equipment failure or improper monitoring conditions
which may endanger the patient being monitored.
CAUTION
Do not locate the monitor in an enclosed area that may
restrict the heat dissipated by it. Any restriction in air
flow causes a rise in internal temperature which may
result in equipment failure.
CAUTION
The monitor must be located no closer than 4 inches (10
cm) from any partition or wall. The monitor should be
approximately 12 inches (30 cm) from any overhead
partition or the ceiling.
Mounting
Recommendations
Marquette Monitoring System Mounting Reference Guide:
¥
Manufacturer recommended methods of mounting the monitor to
various locations.
Software Setup
Section 6: Configuration
¥
Information regarding connection of the monitor to peripherals
Summary of Contents for marquette Dash 2000
Page 12: ...Contents viii DASH 2000 Patient Monitor Revision A 2000 412 001 ...
Page 82: ...Maintenance Repair Log 3 36 DASH 2000 Patient Monitor Revision A 2000 412 001 ...
Page 156: ...UPPER LEVEL ASSEMBLY Spare Parts List 7 10 DASH 2000 Patient Monitor Revision A 2000 412 001 ...
Page 157: ...Revision A DASH 2000 Patient Monitor 8 1 2000 412 001 8 ASSEMBLY DRAWINGS Contents ...
Page 159: ......
Page 173: ......
Page 178: ......
Page 180: ......
Page 181: ......
Page 182: ......
Page 184: ......
Page 196: ......
Page 197: ......
Page 198: ......
Page 200: ......
Page 201: ......