
Maintenance: Checkout Procedures
Revision A
DASH 2000 Patient Monitor
3-17
2000 412-001
Speaker Tests
1.
Set HR limits so, that HR exceeds the limits and alarm tone occurs.
2.
Change the alarm volume of the monitor to 100%.
3.
Verify the speaker volume of the monitor changes accordingly.
4.
Return the volume of the monitor to the level it was previously set to,
before you changed it for this test.
LAN Network Check
Do the following to check the monitoring network:
Check if the ECG and parameter values are displayed correctly on a
Centraloscope or CIC.
Completion
This completes all tests associated with the checkout procedures.
Disconnect the monitor from all test equipment in the following manner:
1.
Set all test equipment power switches to the off position.
2.
Set the monitor front panel power switch to the off position.
3.
Remove all test equipment from the monitor.
Summary of Contents for marquette Dash 2000
Page 12: ...Contents viii DASH 2000 Patient Monitor Revision A 2000 412 001 ...
Page 82: ...Maintenance Repair Log 3 36 DASH 2000 Patient Monitor Revision A 2000 412 001 ...
Page 156: ...UPPER LEVEL ASSEMBLY Spare Parts List 7 10 DASH 2000 Patient Monitor Revision A 2000 412 001 ...
Page 157: ...Revision A DASH 2000 Patient Monitor 8 1 2000 412 001 8 ASSEMBLY DRAWINGS Contents ...
Page 159: ......
Page 173: ......
Page 178: ......
Page 180: ......
Page 181: ......
Page 182: ......
Page 184: ......
Page 196: ......
Page 197: ......
Page 198: ......
Page 200: ......
Page 201: ......